Specifications |
|
| |
Input |
 |
Input Voltage |
|
90-264 VAC |
 |
Input Frequency |
|
47-63 Hz |
 |
Input Current |
|
3.2 A at 115 VAC, 1.6 A at 230 VAC |
 |
Inrush Current |
|
20 A at 115 VAC, 40 A at 230 VAC, cold start +25 °C |
 |
Power Factor |
|
0.98 typical |
 |
Earth Leakage Current |
|
90 µA max at 115 VAC/60 Hz,
150 µA max at 230 VAC/50 Hz |
 |
Input Protection |
|
Internal T4 A/250 V in line and neutral |
| |
Output |
 |
Output Voltage |
|
See datasheet |
 |
Output Voltage Adjustment |
|
Not user-adjustable |
 |
Minimum Load |
|
See table and note 5 of datasheet |
 |
Start Up Delay |
|
2 s max |
 |
Start Up Rise Time |
|
18 ms typical |
 |
Hold Up Time |
|
20 ms minimum at full load & 110 VAC |
 |
Drift |
|
±0.2% |
 |
Line Regulation |
|
±0.5% max |
 |
Load Regulation |
|
See note 7 of datasheet |
 |
Transient Response |
|
4% max deviation, recovery to within 1% in 500 µs for a 25% load change |
 |
Ripple and Noise |
|
2% max pk-pk, 20 MHz bandwidth with a 22 µF electrolytic & a 0.47 µF tantalum capacitor
|
 |
Overvoltage Protection |
|
112-140% Vnom output 1 only, recycle input to reset
|
 |
Overload Protection |
|
110-150% typical, auto recovery |
 |
Overtemperature Protection |
|
Shutdown at +85 °C (singles) or +105 °C (multi outputs) with auto recovery, measured internally |
 |
Short Circuit protection |
|
Trip & restart (hiccup mode) |
 |
Temperature Coefficient |
|
0.04%/°C |
 |
Remote Sense |
|
Single output models only. Compensates for up to 0.5 V drop |
 |
Remote On/Off |
|
On = Logic low or open circuit,
Off = Logic high or short circuit |
 |
Initial Set Accuracy |
|
±2%
|
| |
General |
 |
Efficiency |
|
80% maximum |
 |
Isolation |
|
4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
|
 |
Switching Frequency |
|
100 kHz ±12 kHz |
 |
Power Density |
|
4.76 W/In3 |
 |
Power Fail Detect |
|
AC OK = TTL logic high
AC not OK = TTL logic low |
 |
MTBF |
|
350 kHrs to MIL-HDBK-217F at +25 °C, GB
|
 |
Signals |
|
Power fail detect, remote on/off |
| |
Environmental |
 |
Operating Temperature |
|
0 °C to +70 °C, derate linearly from 100% power at +50 °C to 50% power at +70 °C |
 |
Storage Temperature |
|
-40 °C to +85 °C |
 |
Cooling |
|
‘C’ versions have internal fan,
‘B’ versions are derated with convection cooling or require 25 CFM (see note 4 of datasheet) |
 |
Operating Humidity |
|
Up to 95% RH, non condensing |
 |
Operating Altitude |
|
2500 m |
 |
Shock |
|
Half sine pulse. acceleration 50 g, 11 ms duration, 3 shocks in each direction, unpackaged, non-operating
|
 |
Vibration |
|
Sinusoidal 10-55 Hz, amplitude 0.15 mm, 30 min/axis, unpackaged, non-operating
|
| |
EMC and Safety |
 |
Emissions |
|
EN55011 & FCC Class B conducted & radiated |
 |
Harmonic Currents |
|
EN61000-3-2, Class A |
 |
Voltage Flicker |
|
EN61000-3-3 |
 |
ESD Immunity |
|
EN41000-4-2, level 3 Perf criteria A |
 |
Radiated Immunity |
|
EN61000-4-3, level 3 Perf criteria A |
 |
Conducted Immunity |
|
EN61000-4-6, 3 V Perf Criteria A |
 |
EFT/Burst |
|
EN61000-4-4, level 3 Perf criteria A |
 |
Surge |
|
EN61000-4-5, level 3 Perf criteria A |
 |
Dips & Interruptions |
|
EN61000-4-11, 30% 10 ms,
60% 100 ms, >95% 5000 ms,
Perf Criteria A, B, B |
 |
Safety Approvals |
|
EN60601-1, UL60601-1,
CSA22.2 No. 601-1 per cUL |